Scientific Paper
Electrophoresis and Quartz Crystal Microbalance Instrumentation to Sense Nanoplastics in Water
Summary
Nanoplastics in water can be pulled onto a QCM electrode by electrophoresis and weighed as a drop in resonance frequency. Masking most of the electrode so that the particles collect on a small central area raises the sensitivity; the method told polyethylene-spiked bottled and mineral water apart from clean samples, with Raman spectroscopy and microscopy confirming where the particles settled.
Summary written by the openQCM team. The original abstract is on the publisher’s page.