openQCM test of quartz crystal in contact with liquid

The entire frequency data set of quartz crystal in contact with air and with pure water using two different openQCM instruments

Table of Contents

Here we report in detail the verification test of openQCM Quartz Crystal Microbalance in contact with pure water according to the theory based on Kanazawa – Gordon equation

THEORY

Kanazawa and Gordon derives a simple relationship [1][2], which expresses the frequency variation of a quartz crystal in contact with a liquid in terms of material parameters of the fluid and the quartz. The relationship is:

 

\Delta f = - f_0^{3/2} \cdot \left( \frac{\eta_L \rho_L}{\pi \mu_Q \rho_Q} \right)^{1/2}

 

where:

 

 

  f_0 nominal vibration frequency of the dry quartz crystal
\eta_L absolute viscosity of the liquid
\rho_L density of the liquid
\mu_Q shear modulus of the quartz crystal
\rho_Q density of the quartz crystal

 

 

In other words, the frequency variation of a quartz crystal in contact with liquid depends on the viscosity – density product of the liquid under investigation. The Kanazawa – Gordon equation is verified under the following assumption:

i) quartz crystal is a perfectly elastic solid, which means energy dissipation due to crystal internal friction and mounting friction are neglected
ii) the liquid is Newtonian, which means it is a purely viscous liquid,
iii) the liquid can be approximated to a semi-infinite layer, which means the penetration depth of the shear wave inside the liquid is very much smaller than the liquid thickness in contact with the quartz crystal.

MATERIAL and METHODS

Two different openQCM instruments were compared in order to perform the verification test under liquid. It was used a 10 MHz AT-cut quartz crystal sensors with 6 mm diameter gold electrode. Pure water was injected in the openQCM measurement chamber using a syringe pump. The measurement chamber volume is roughly 30 µL (micro liters). The material parameters of the fluid and the quartz are:

 

f_0 \simeq 10 \,  MHz
\eta_L = 1.002 \times 10^{-2}  \, g \cdot cm^{-1} \cdot s^{-1} (T = 20^\circ )
\rho_L = 0.9982 \:  g \cdot cm^{-3} (T = 20^\circ )
\mu_Q = 2.947  \times 10^{11}\, g \cdot cm^{-1} \cdot s^{-2}
\rho_Q = 2.648  \, g \cdot cm{-3}

 

According to Kanazawa Gordon equation the frequency variation from air to pure water is:

 

\Delta f = 2020 \, Hz

 

The frequency shift caused by injection of pure water in the measurement chamber was measured using the following procedure:

  1. Start openQCM and wait for the steady state, qualitatively when temperature reaches equilibrium.
  2. Measure and record the openQCM initial frequency over a time interval of about 30 minutes. The initial frequency corresponds to quartz crystal in contact with air.
  3. Inject pure water in the measurement chamber using the syringe pump.
  4. Wait for the new steady state, qualitatively when temperature reaches a new equilibrium typically at higher temperature.
  5. Measure and record the openQCM final frequency over a time interval of about 30 minutes. The final frequency corresponds to the quartz crystal in contact with pure water.
  6. Calculate the frequency difference = final frequency – initial frequency.

TIP and TRICKS

Here some tip and tricks for using openQCM in contact with liquid.

1. Holder and quartz crystal connection: pay attention to mount the quartz crystal into the openQCM holder (model HC-48/U) as shown in Figure 1

How to properly mount the quartz crystal sensor into the HC-48/U holder.
Figure 1 How to properly mount the quartz crystal sensor into the HC-48/U holder.

2. Bubble: please make sure no bubbles are present in the measurement chamber during your entire test. We have designed a transparent measurement cell exactly for this reason.

3. Liquid injection: in order to avoid bubble it is a good practice to inject the liquid in the measurement chamber, using a syringe pump, a peristaltic pump or whatever, as shown in Figure 2. You should use the larger hole as inlet (IN) and the smaller one as outlet (OUT).

Liquid inlet and outlet of the openQCM measurement cell, which correspond to the larger hole and the smaller hole respectively
Figure 2 Liquid inlet and outlet of the openQCM measurement cell, which correspond to the larger hole and the smaller hole respectively

4. o-ring mechanical stress: sometimes the o-ring can become more rigid when it is in contact with liquid, in particular for the first time. It causes a long-term drift in the frequency behavior. In this case, the mechanical stresses can be released by knocking hard with one finger on the measurement chamber.

5. Alignment: try to align at your best the quartz crystal electrode to the centre of the measurement cell. In this way the quartz crystal will be uniformly in contact with the liquid

EXPERIMENTAL SECTION

The experimental results of the entire test for both openQCM instruments are shown in Figure 3. The entire test verification takes approximately 2 hours.

The entire frequency data set of quartz crystal in contact with air and with pure water using two different openQCM instruments
Figure 3 The entire frequency data set of quartz crystal in contact with air and with pure water using two different openQCM instruments

 

openQCM instrument #01

The initial frequency corresponds to the quartz crystal in contact with air. The initial frequency was measured on a time interval of about 30 minutes at temperature of about 23.5 °C. After the injection of pure water, it is necessary to wait for the new steady state. The final frequency corresponds to the quartz crystal resonator in contact with pure water. The final frequency was measured on a time interval of about 30 minutes at temperature of about 26.5 °C. The table below summarizes the experimental results for openQCM #01 instrument and the Figure 4 shows the frequency data set in air and liquid environment

 

 

initial frequency (air) f_{in} = 10026247.3 ± 0.2 Hz
final frequency (water) f_{fin}= 10024160.8 ± 0.7 Hz
frequency difference \Delta f = 2086.5 ± 0.7 Hz

 

Table 1 Experimental results for openQCM #01 device: initial frequency corresponds to quartz crystal in contact with air, final frequency corresponds to quartz crystal in contact with liquid pure water

 

 

openQCM #01 detail of the frequency data set corresponding to the quartz crystal in contact with air and liquid pure water. In both cases, the averaging time is about 30 minutes
Figure 4 openQCM #01 detail of the frequency data set corresponding to the quartz crystal in contact with air and liquid pure water. In both cases, the averaging time is about 30 minutes

 

openQCM instrument #02

The same procedure was used for the second openQCM device. The initial frequency was measured on a time interval of about 30 minutes at temperature of about 24.0°C The final frequency  was measured on a time interval of about 30 minutes at temperature of about 26.2 °C. The Table 2 below summarizes the experimental results for openQCM #02 instrument and Figure 5 shows the frequency data set in air and liquid environment

 

 

initial frequency (air)   f_{in} = 10020185.6 ± 0.2 Hz
final frequency (water)   f_{fin}= 10018090.4  ± 0.4 Hz
frequency difference   \Delta f = 2095.2  ± 0.5 Hz

 

Table 2 Experimental results for openQCM #02 device: initial frequency corresponds to quartz crystal in contact with air, final frequency corresponds to quartz crystal in contact with liquid pure water

 

 

openQCM #02 detail of the frequency data set corresponding to the quartz crystal in contact with air and liquid pure water. In both cases, the averaging time is about 30 minutes
Figure 5 openQCM #02 detail of the frequency data set corresponding to the quartz crystal in contact with air and liquid pure water. In both cases, the averaging time is about 30 minutes

 

RESULTS and CONCLUSION

The experimental results demonstrate a good agreement between the frequency shift measured by openQCM to the passage from air to pure water and the theoretical prediction of the Kanazawa Gordon equation. The theory predicts a frequency shift of about 2kHz for a 10 MHz quartz crystal resonator from air to pure water at ambient temperature. The frequency shift measured using openQCM is very close to the theoretical value. It was also demonstrated the openQCM measurement reproducibility by testing and comparing two different openQCM instruments under the same experimental conditions. The test and results we have reported shown a very simple and practical way to verify the working condition of openQCM and it is also a good way to make some practice with quartz crystal microbalance in contact with liquid.

 

FURTHER READINGS

Kanazawa and Gordon published their theoretical and experimental results in two papers in the ’80s. The papers are pioneering and inspiring masterpiece in this field. You should read them!

[1] K Keiji Kanazawa and Joseph G Gordon II, “Frequency of a quartz microbalance in contact with liquid”, Anal. Chem., 1985, 57 (8), pp 1770–1771
doi: 10.1021/ac00285a062

[2] K Keiji Kanazawa and Joseph G Gordon II, “The oscillation frequency of a quartz resonator in contact with liquid”, Analytica Chimica Acta, Volume 175, 1985, Pages 99-105
doi:10.1016/S0003-2670(00)82721-X

enjoy,
marco

Share:

Facebook
Twitter
LinkedIn

8 Responses

  1. Thanks Marco for the sharing. My experience using openQCM is that it did require some time (~30 minutes) to get the temperature stable for frequency measurement. But I found if we connected the USB cable before measurement for 30-60 minutes without opening up the software, the frequency signal was easy to stablize within a few minutes.
    Jason

  2. Hi Jason, you are very welcome !
    openQCM needs roughly 30-60 min for thermal stability depending on your environment condition.
    It mainly depends on the warming up time of the electronic board. So your procedure is totally right.
    It is a good practice to wait for the temperature stability before starting the QCM frequency measurements.

  3. Hii all, Does anyone get the problems when using QCM in contact with a large volume of liquid. In my case, I design chamber with diameter ~ 10.8mm and heigh ~2mm. I using Arduino shield of openQCM. When I using syringe pump to flow in the PBS in chamber with flow rate is 6uL/s. First, the frequency of QCM ~ nearly 5MHz, but when the chamber was filled in with PBS, the frequency is 8MHz, or 9Mhz, sometime 15Mhz. Actually, I can not understand why it is ?

    Is there any problem in the Pierce oscillating circuit ?

    1. Hi Nhan,
      Generally is a goood approach to use a minimal chamber height. As matter of fact the QCM will be not sensitive to the volume above few microns, but most probably in this way you will introduce viscous effect of the “large” liquid mass. Furthermore you shoud control the right quartz mounting (as explained in the label attached to your package). If you will mount the quartz in a wrong way, you will observe high instability and big frequency gapps (generally only in liquid environment). Relatively your last question, our shield is calibrate for 10 MHz sensors. As matter of fact, despite the oscillator can work with all frequencies between few KHz p to 28 MHZ, the use at differennt frequencies need to change the 2 capacitors C1 and C2.
      Furthermore, pay attention if you use arduino micro. The micro has a clock at 16MHz. As a consequency you will not be able to use some frequency near to half freq clock (eg.: 7Mhz , 8 MHz, 9 MHz etc). You can solve this last point by using a Teesy board instead of an Arduino Micro.

  4. Hi openQCM.
    I had review your circuit shield. And this circuit was based on Pierce oscillating circuit. So as far as I know and also some papers related to design the electronic circuit for QCM. When QCM is in liquid, due to the viscosity of solution, and “soft-layer” between antigent-antibody so the frequency of QCM will be damped or dissipated in respect to time.
    So, the problems is that ” How long the period of time does openQCM be stable and reliable operation ?” Because in many commercial oscillating circuits have a AGC module ( Automatic Gain Control) to remain the amplitude of oscillation of QCM, so QCM can be operated for a long time. This is my big question about electronic circuit operation of QCM. Because many papers I had read, the autors usually use QCM sensor for real- time monitoring from immobilization antibodies to detection of antigen process. Generally, these experiment took a long time, ~ 24h or more. the OpenQCM-Dissipation is going to launch, and this is a best solution to use QCM in liquid, I believe that.

    Best regards and warmly to wait openQCM-Dissopation version

    Nhan ND, from UNIST, Korea of Republic

  5. Hi openQCM.
    I had review your circuit shield. And this circuit was based on Pierce oscillating circuit. So as far as I know and also some papers related to design the electronic circuit for QCM. When QCM is in liquid, due to the viscosity of solution, and “soft-layer” between antigent-antibody so the frequency of QCM will be damped or dissipated in respect to time.
    So, the problems is that ” How long the period of time can openQCM be stable and reliable operation ?” Because in many commercial oscillating circuits have a AGC module ( Automatic Gain Control) to remain the amplitude of oscillation of QCM, so QCM can be operated for a long time. This is my big question about electronic circuit operation of QCM. Because many papers I had read, the autors usually use QCM sensor for real- time monitoring from immobilization antibodies to detection of antigen process. Generally, these experiment took a long time, ~ 24h or more. the OpenQCM-Dissipation is going to launch, and this is a best solution to use QCM in liquid, I believe that.

    Best regards and warmly to wait openQCM-Dissopation version

    Nhan ND, from UNIST.

Leave a Reply

Your email address will not be published. Required fields are marked *

On Key

Related Posts

Quartz Crystal Microbalance with active temperature peltier control openQCM Wi3

An exciting year

Hello, everyone. This 2018 has been a very exciting year. The openQCM project is growing beyond our expectations. We have launched 2 new devices and

openQCM – Powered by Novaetech S.r.l

COOKIES & PRIVACY POLICY