Scientific Paper
MoS2/MXene bilayer quartz crystal microbalance sensor with enhanced sensitivity and selectivity for room-temperature xylene detection
Summary
Xylene sensing at room temperature with QCMs coated with MoS2, MXene and stacked MoS2/MXene films. The bilayer worked best, reaching 2.46 Hz per ppm and a 1.23 ppm detection limit over 2.5–600 ppm, which the authors attribute to charge transfer at the MoS2/MXene junction; humidity lowers the response, so it must be controlled or compensated.
Summary written by the openQCM team. The original abstract is on the publisher’s page.
Related work.
- MULTIPLEXED BIOSENSOR USING QUARTZ-ON-SILICON MICRO-ACOUSTIC (QSIM) TECHNOLOGY FOR IN-VITRO LABEL-FREE INVESTIGATION OF HEMOSTASIS 2023
- Luminescent polysaccharides thin films with potential bactericidal activity 2022
- An approach to compare performance of surgical masks for fighting against the COVID-19 pandemic 2022
- Study of Multichannel QCM Prospects in VOC Detection 2021
- Χαρακτηρισμός νανοσωματιδίων προερχομένων από διεργασίες καύσης και συνθετικών νανοσωματιδίων και αξιολόγηση επιπτώσεων τεχνολογιών ελέγχου 2020
- Příprava aminových plazmových polymerů a jejich interakce s ECM proteiny 2019
New openQCM papers, by email.
A selection of new publications measured with openQCM, with software releases and application notes. A few emails a year.
Browse all papers.
Hundreds of peer-reviewed publications cite openQCM. Search full-text and filter by instrument, year, journal or topic.