Scientific Paper
High-performance formic acid gas detection using a quartz crystal microbalance functionalized with Cdot-decorated polyvinyl acetate nanofibers
Summary
A formic acid vapour sensor made from a QCM coated with electrospun polyvinyl acetate nanofibres and carbon dots. The carbon dots raised the sensitivity about twentyfold, from 0.126 to 2.552 Hz per ppm, with response and recovery in 27 s and 63 s, good selectivity and stable behaviour over 29 days of tests.
Summary written by the openQCM team. The original abstract is on the publisher’s page.
Related work.
- Aptamer-molecularly imprinted polymer sensors for the detection of bacteria in water 2025
- Polymère à empreinte moléculaire pour la reconnaissance et l’immobilisation des bactéries photosynthétiques 2024
- MULTIPLEXED BIOSENSOR USING QUARTZ-ON-SILICON MICRO-ACOUSTIC (QSIM) TECHNOLOGY FOR IN-VITRO LABEL-FREE INVESTIGATION OF HEMOSTASIS 2023
- Luminescent polysaccharides thin films with potential bactericidal activity 2022
- An approach to compare performance of surgical masks for fighting against the COVID-19 pandemic 2022
- Study of Multichannel QCM Prospects in VOC Detection 2021
New openQCM papers, by email.
A selection of new publications measured with openQCM, with software releases and application notes. A few emails a year.
Browse all papers.
Hundreds of peer-reviewed publications cite openQCM. Search full-text and filter by instrument, year, journal or topic.