openQCM Store Talk with us
← The instrument familyHardware monograph / No. 06openQCM Holder
Sensor holder · all FR-4 · from the bench to vacuum

openQCM Holder · Vacuum-compatible QCM holderThe board is the housing.

openQCM Holder has no case. Three layers of FR-4, the glass-epoxy of circuit boards, are at once the structure, the crystal seat and the wiring. Under 5 grams and 4 mm thin, with no plastic housing, it carries a QCM crystal from the open bench into a vacuum chamber.

Fig. 1 · Three layers and a crystalDrag to open the stack
Bottom · pads or plug Middle · crystal seat, sensor Quartz crystal · 14 mm Top · window on the electrode
weight
< 5g
thickness
4mm
without the plug
25 × 18.6mm
crystals, 14 mm
1–10MHz
The openQCM Holder, USB version, next to a one-euro coin
USB version, next to a one-euro coin
Three layers

Electronics as structure.

Three FR-4 layers clamp a 14 mm wrapped-electrode crystal (1 to 10 MHz) and carry its contacts, an MCP9808 temperature sensor on I²C (−40 to +125 °C) and a Pierce oscillator whose output is a TTL signal at the crystal's resonance frequency. Building the holder from circuit board removes the plastic housing, and with it a source of mechanical play.

Two ways to connect

A plug, or seven gold pads.

The USB version plugs into your own acquisition electronics through a USB 3.0 connector with its own pinout. The pad version is made for soldering: seven gold pads take the wires of a vacuum feedthrough, so only the holder sits inside the chamber.

Not a USB device

Never plug the USB version into a computer. The connector is a USB 3.0 plug, but its pins carry the crystal, the oscillator output and the I²C bus, not USB: a standard port can damage the holder.

PinSignalUse
1SDAI²C data, temperature sensor
2QCM OUTCrystal
3QCM INCrystal
4NCNot connected
5SIG OUTPUTOscillator output
6GNDGround
7V+Supply
8GNDGround
9SCLI²C clock, temperature sensor

Same signals, different order: on the plug the crystal is on pins 2–3, on the pads on 2 and 6. Check twice before you solder.

Vacuum

What the manual says, and what it does not.

All-FR-4 construction makes the holder suitable for atmospheric and vacuum environments. The manual gives no pressure range, outgassing figure or bake-out temperature: check them for your process, and keep the rest of the electronics outside the chamber.

Vacuum chamber Holder · 7 pads Feedthrough Your readingelectronics inside: FR-4 and the crystal only
Specifications

The data sheet.

From the official manual. Where two documents disagree, the figure is left out until it is confirmed.

Sensor

Crystal
Single, 1–10 MHz, 14 mm, wrapped electrode
Electronics
Pierce oscillator, TTL output
Quantity
Frequency
Temperature
MCP9808 on I²C, −40 to +125 °C

Mechanics

Material
FR-4, three PCB layers
Size
25 × 18.6 × 4 mm; 40.2 mm long with the plug
Weight
< 5 g

Electrical

Connection
USB 3.0 plug (custom pinout) or gold solder pads
Supply
3.3–5 V
Current
Max 80 mA at 5 V
Talk with us

Put a QCM in the chamber.

Tell us about your vacuum rig or custom setup, and we will help you with crystals, wiring and reading electronics.

Colophon

Explore. Measure.
Publish. Repeat.

openQCM is a brand of Novaetech S.r.l. · Open hardware since 2014 · © 2026 · Cookies & Privacy Policy · Images CC BY-NC-SA 4.0, commercial use: ask us